    Liquid crystal type metal tube rotameter

    18金屬筦轉子(zi)流量計(液晶型).jpg
    • 18金屬筦轉子流量(liang)計(液晶型).jpg
    ZW-LZ series Liquid crystal type metal tube rotameter Is composed of a measuring tube and a local in…

    Description:

    Productoverview:

    ZW-LZ series Liquid crystal type metal tube rotameter Is composed of a measuring tube and a local indicator. Local indicator can also choose to install additional electrical components, is suitable for measurement of liquid and gas. The principle of variable area measurement, measuring unit consists of a conical measuring tube free to move up and down, change in the pipeline flow area, with vertical position changes the size of the float flow in the measuring tube is changed, the float position is transferred to the dial indicator on the magnetic transmission system. The flow is indicated the magnitude of the output signal can realize remote control and display.

    Second, working principle:

    ZW-LZ Series LCD type metal tube rotameter is the medium to be measured through a conical measuring tube float from upper and lower ends of the differential pressure generated form the upward force, when the float was immersed in a fluid on the lift is greater than the weight of the float, the float will rise, annulus area increases, the annulus fluid velocity decreased immediately float the lower end of the differential pressure is reduced, to float upward force also decreases with, until the lift is equal to the weight of the float when immersed in a fluid, the float will stabilize at a certain height. The float position that corresponds to the size of the measured medium flow. A built-in magnet, the float with the medium moves up and down, with the movement of the float magnetic field changes.

    A, for in situ, by rotary magnetic float and the local indicator in the inner magnetic coupling, rotate, and drives the pointer, through the dial indicates the size of the flow.

    B, the intelligent, by a solid state magnetic sensor intelligent indicator in the magnetic field will change into electrical signal output by A/D transform, D/A microprocessor, LCD liquid crystal display to show the size and cumulative flow flow.

    Third, application:

    ZW-LZ Series LCD type metal tube rotameter is reliable, low maintenance, long service life; widely used in chemical industry, power generation, food, water treatment etc.. The complex and harsh environmental conditions, the flow measurement process and various medium conditions in.

    Fourth, the main features:

    1, all metal structure design, strong, simple, high pressure resistance, high temperature resistance, corrosion resistance, long service life;

    2, a short trip, the total height of 250mm, easy to install;

    3, mechanical pointer instructions LCD liquid crystal display instantaneous flow, instantaneous flow, cumulative flow;

    4, intelligent indicator, using foreign Z new signal processing chip, modular design, no hysteresis;

    5, data recovery, data backup, power down protection and fault self diagnosis function;

    6, HART type indicator, two-wire 24VDC power supply, 4-20mA standard current signal output of HART protocol stack;

    7, through the HART communication manipulator configuration and parameters can be adjusted;

    8, with the PC serial connection, through the Windows program, to achieve on-site calibration of the instrument;

    9, can be used in flammable and explosive occasions;

    10, vertical and horizontal, upper lower, bottom side, side into the side such as the installation of a variety of forms, flange or threaded connection.

    Fifth, technical parameters:

    1, measurement range: water 2.5 ~ 100000l/h (20 C)

    2, range: 10:1

    3, 1.0, 1.5, precision grade: 2.5

    4, working pressure: DN15 DN25, DN50, PN4.0MPazui, DN80, DN100 10.0MPa PN1.6MPazui 6.4MPa

    5, medium temperature: -40 to 300 DEG C

    6, medium viscosity: DN15: <5mPa - s eta (F15.1 ~ F15.3)

    <30mPa - S (F15.4 - F15.8)

    DN25: <250mPa - s.

    DN50: <300mPa - S - DN150.

    7, the ambient temperature: LCD type: -40 to 85 DEG C

    Pointer type: -40 to 120 DEG C

    8, connection: flange (executive standard DIN2501 or as required by standard flange manufacturing)

    9 meter height: 250mm

    10, cable interface: M20*1.5

    11, power supply: 24VDC two-wire 4 ~ 20mA or 85 ~ 265VAC 50/60Hz (remote type)

    12, alarm output: alarm relay output lower or upper limit of instantaneous flow (contact capacity Zui 5A-250VAC) or open collector output (internal impedance Zui 100mA-30VDC 100 ohm).

    13, the cumulative output pulse: pulse output, Zui interval of every 10 seconds a pulse (AC) or a pulse every 50 milliseconds.

    14, double liquid crystal display: LCD display, display the instantaneous flow and cumulative flow.

    15: Exia II CT4, explosion-proof.

    16, the measuring tube material: 316 stainless steel (common type) or Teflon liner (corrosion type).

    Six, product maintenance:

    1, in the process of long-term use, the pipeline inevitably there will be ferromagnetic impurity adsorption on the float, it will float stuck or affect the measurement accuracy, from time to time to clean the sensor float; if the installation of magnetic filters should be cleaned regularly.

    2, the instrument indicator with electronic components, instrument wiring or demolition of housing, to housing screws, must ensure sealing, prevent impurities, or other material into the water, and also to ensure reliable earthing of the meter casing.

    3, the instrument is installed, in use, should pay attention to:

    When the valve is opened to avoid the formation of pressure shock makes the float slamming limiting device, causing damage to the instrument, slowly open the valve!

    The measurement of gas meter is equipped with gas damper, Zui to decrease the float concussion; to ensure the stability of the float, a throttle valve can be installed in the instrument export.

    4, for the remote type instrument, we must first ensure that the instrument wiring correctly, keeps power; for dangerous situations, must choose the explosion-proof, and the installation and use of proof required by the.

    Troubleshooting seven fault:

    1, check the alarm set value size SU.

    2, check the deviation set D value can not be too large.

    3, FUN function, logic function is correct. HA-A said the upper limit is logical. LA-A said the lower limit is.

    4, circuit board failure, replace the board.

    5, if the lcd barcode indicates the correct output, no action, the cathode can check the external power and external power supply is connected with the anode of instrument power supply.

    If the metal tube rotameter jitter in use, according to the jitter amplitude to determine what kind of amplitude jitter.Divided into mild and moderate jitter jitter jitter and jitter.

    1, a slight jitter due to fluctuations caused by the media. Can be used to increase the damping of the way to overcome.

    2, moderate jitter caused by medium flow state: general. For general gas as the medium pressure instability caused by.The regulator or stabilizer to overcome or increase the damping of gas meter.

    3, severe jitter: mainly due to medium gas pulsation, operation pressure instability or give users pressure, temperature, flow meter and the actual state does not have large differences over range instrument.

    Eight, installation requirements.

    Remove the filler, 1 meter box after check, the meter should be installed vertically in the pipeline without vibration, such as to the level of the installation, the order should be additional instructions. The angle between the line and the plumb line flowmeter center should not exceed 5 degrees.

    In 2, the newly installed pipeline flow meter should be installed before the new pipeline clean metal pipe, if the measured medium containing impurity particles or bubbles, you should install filtering device or exhaust in flowmeter installation, the fluid must be flowing from down to up, if the fluid flow from top to bottom, then the flowmeter can not work. If the measured medium is pulsating flow, it should be appropriate to the size of the buffer device is arranged in the downstream, such as surge tank to eliminate pulsation.

    3, in order to check the meter zero and disassembly and maintenance and debugging, should stop valve and bypass valve installed in the upstream and downstream flow meter.

    4, an instrument according to the provisions after installation, should close the flow meter on the downstream shut-off valve, check again with the two meter instrument wiring, error free, open the two meter power supply is the instantaneous flow rate should be zero. For normal use, you should first open the upstream valve is fully open, with a metal tube flowmeter downstream from the valve slowly adjust the flow, instantaneous flow should change. Stop working when they should close the valve, then close the upstream downstream of the valve.

    Nine, the measuring range and the pressure loss:

    Decay path

    Common path

    measuring range

    Zui high pressure loss

    DN (mm)

    DN (mm)

    The air (m3/h)

    Water (L/h)

    The air (KPa)

    Water (KPa)

    Fifteen

    Fifteen

    0.05-0.5

    1.6-16

    Seven

    Six point four

    0.07-0.7

    2.5-25

    Seven point one

    Six point five

    0.11-1.1

    4.0-40

    Seven point two

    Six point five

    0.18-1.8

    6.0-60

    Seven point three

    Six point six

    0.28-2.8

    10-100

    Seven point five

    Six point six

    0.40-4.0

    16-160

    Eight

    Six point eight

    0.70-7.0

    25-250

    Ten point eight

    Seven point two

    1.00-10

    40-400

    Ten

    Eight point six

    1.60-16

    60-600

    Fourteen

    Eleven point one

    Twenty-five

    -

    80-800

    -

    Fourteen

    Twenty-five

    3.00-30

    100-1000

    Seven point seven

    Seven

    4.50-45

    160-1600

    Eight point eight

    Eight

    7.00-70

    250-2500

    Twelve

    Ten point eight

    11-110

    400-4000

    Nineteen

    Fifteen point eight

    Fifty

    -

    500-5000

    -

    Sixteen point two

    Fifty

    18-180

    600-6000

    Eight point six

    Eight point one

    25-250

    1000-10000

    Ten point four

    Eleven

    Eighty

    40-400

    1600-16000

    Twelve point six

    Seventeen

    Eighty

    50-500

    2000-20000

    Fifteen point five

    Six point three

    One hundred

    75-750

    2500-25000

    Seventeen Point Two

    Eight point one

    One hundred

    100-1000

    4000-40000

    -

    Nine point five

    One hundred and twenty-five

    160-1600

    6000-60000

    -

    Ten

    One hundred and twenty-five

    250-2500

    8000-80000

    -

    Sixteen

    One hundred and fifty

    -

    10000-100000

    -

    Twenty


    300-3000

    12-120m3/h

    -

    Twenty


    400-4000

    15-150m3/h

    -

    Twenty-five

    10, encoding selection:

    Specification code

    Encoding instructions

    The name of the factory

    ZW

    Zhongwei control

    Instrument types

    LZ

    Metal tube rotameter

    structural style

    /

    In the next

    H1

    The left into the right

    H2

    Right into the left out

    AA

    The side into the side

    LA

    The bottom side

    S

    Threaded connection

    M

    Clamp connection (sanitary polishing tube)

    Size code

    - XXX

    Cases: Fifteen Express DN15

    The measuring tube material

    R1

    1Cr18Ni9Ti , Three hundred and four

    R0

    0cr18Ni12M02Ti , Three hundred and sixteen

    RL

    316L

    RP

    (PTFE F-lined )

    Ni

    (Hasloy Hastelloy C)

    Ti

    (TitHXium titanium alloy )

    Media type

    L

    liquid

    G

    Gas

      working temperature 

    H1

    - 40 C -100 C

    H2

    - 80 C -250 C

    Rated pressure

    One point six

    1.6MPa

    Four

    4.0MPa

    Sixteen

    16MPa

    Twenty-five

    25MPa

    Additional structure

    O

    nothing

    Z

    Damping type

    T

    Jacket

    H

    High temperature type

    Y

    High pressure type

    Form indicator

    M1

    Local indication

    M2

    Intelligent

    M3

    Intelligent ( Flameproof )

    The output signal in the communication

    N

    nothing

    A

    4-20mA

    H

    4-20mA+Hart

    Alarm controller

    K1

    The upper limit alarm

    K2

    Alarm

    K12

    The upper and lower limit alarm

    Explosion proof type

    E1

    Intrinsically safe EX IaCT5 Square shell

    E2

    Flameproof EXiibt4 Park Housing

    Power supply

    Zero

    24VDC (can be omitted )

    Two

    220VAC

    Three

    Three point six Volt lithium battery

    Eleven, design selection and ordering information:

    1, please provide the pipe size, wall thickness, material requirements and accuracy requirements;

    2, please provide a liquid or gas, such as the name, commonly used commonly used temperature, working pressure, common flow, density, viscosity etc.;

    3, choose anti-corrosion type, should indicate the measured corrosion and concentration etc.;

    4, if there are special to show, please contact our sales department.

    Customer service service commitment:

    1, our series of metal tube flowmeter series product quality problems within twelve months, responsible for free maintenance.

    2, the warranty period if the products of our company have any quality problems, our company is responsible for free repair or replacement.

    3, the warranty period if the user use undeserved, cause damage to the product, our company is responsible for the maintenance, collect the damaged parts cost.

    4, after the warranty period of product quality problems, our company is responsible for the maintenance, collect the damaged parts cost.

    5, if the user needs to our service, our company in the province of the user service personnel arrived at the scene within 12 hours, arrived at the scene 48 hours outside users.