    Type Selection Instructions for Micro Flow Metal Tube Rotor Flowmeter
    • File Name:Type Selection Instructions for Micro Flow Metal Tube Rotor Flowmeter
    • Add Time:2019-04-12 10:30:20
    • Document Size:
    • Recommendation:
    • Immediately Download:
    Download Instructions:

    Productoverview:
    ZW-LZZW series of micro flow of metal tube rotameter has the characteristics of strong, stable, reliable and wide application range, can be used for measuring gas and liquid.All metal design to ensure its high pressure, high temperature stability, can withstand extreme environmental conditions.The position of the magnetic induction transmitted to an analog indicator displays the measured values.Without power.
    Two, the structure and principle of measurement:
    1, structure:
    ZW-LZZW series of micro flow of metal tube rotameter is mainly composed of three parts:
    A (Intelligent indicator, indicator, local indicator);
    B, float;
    C, a conical measuring chamber.
    2, the principle of measurement:
    On the lower end of the measured medium from the measuring tube through a conical float generating differential pressure form the upward force, when the float on the lift is greater than immersion float in the fluid weight, the float will rise, annulus area increases, the annulus fluid velocity decreased immediately, float on the lower pressure reduced, to float up with the decrease in stress, until the lift is equal to the weight of the float when immersed in a fluid, the float will stabilize at a certain height.The float position that corresponds to the size of the measured medium flow.A built-in magnet, the float with the medium moves up and down, with the movement of the float magnetic field changes.
    A, for in situ, by rotary magnetic float and the local indicator in the inner magnetic coupling, rotate, and drives the pointer, through the dial indicates the size of the flow.
    B, the intelligent, by a solid state magnetic sensor intelligent indicator in the magnetic field will change into electrical signal output by A/D transform, D/A microprocessor, LCD liquid crystal display to show the size and cumulative flow flow.
    Three, widely used:
    ZW-LZZW series of micro flow of metal tube rotameter is widely used in petroleum, chemical, power, pharmaceutical, food, water treatment, etc..The complex, harsh environmental conditions, the flow measurement process and various medium conditions in.
    Fourth, the main features:
    1, all metal structure, strong stability, suitable for high temperature and high pressure, strong corrosive medium.
    2, using the isolation design, imported relay output, to ensure reliable operation.
    3, in all sectors of the measurement of gas and liquid, the measuring part can be using different materials to adapt to various medium conditions.
    4, the vertical and horizontal installation, more suitable for different occasions.
    5, the magnetic coupling structure to ensure data transmission, more stable signal.
    6, magnetic filter, with thermal insulation or jacket design.
    7, the vertical and horizontal installation, more suitable for different occasions.
    8, can be used in flammable and explosive occasions.
    Five, technical parameters:
    1. measuring range: water (20 C) (0.3 ~ 100) L/h; air (20 degrees 101325Pa) (1.2 ~ 3400) L/h;
    2. range: 10:1;
    3. accuracy level: 4
    Maximum working pressure: 4. 2.5MPa (according to the need of increase)
    5. medium temperature: -80 ~ +150 C
    6. environment temperature: -25 ~ +65 C
    7. connection: 1/4 "NPT thread or phi 6 (phi 8) sleeve connection / vertical connection according to the level of user requirements
    8. limit alarm: 1~2 alarm switch (optional)
    9. dimensions: overall size: 110mm; installation center distance: 90mm;
    Sixth, product maintenance:
    1, in the process of long-term use, the pipeline inevitably there will be ferromagnetic impurity adsorption on the float, it will float stuck or affect the measurement accuracy, from time to time to clean the sensor float; if the installation of magnetic filters should be cleaned regularly.
    2, instrument indicator equipped with electronic components, instrument wiring or demolition of housing, to housing screws, must ensure sealing, prevent impurities, or other material into the water, and also to ensure reliable earthing of the meter casing.
    3, instrument installation, on first use, should pay attention to:
    When the valve is opened to avoid the formation of pressure shock makes the float slamming limiting device, causing damage to the instrument, slowly open the valve!
    The measurement of gas meter is equipped with gas damper, to minimize the shock; to ensure the stability of the float, a throttle valve can be installed in the instrument export.
    4, for the remote type instrument, we must first ensure that the instrument wiring correctly, keeps power; for dangerous situations, must choose the explosion-proof, and the installation and use of proof required by the.
    Seven, the installation should follow the following steps:
    1, before installing the instrument to process piping, should remove all installed and checked for any transport damage;
    2, in order to prevent in transit, with a rubber band fixed pointer indicator, indicator should open the left cover it gently removed before connection, do not change the pointer position;
    3, open the indicator of the right cover, cable through the cable introducing device, according to the wiring diagram wiring rotation transmitter, pressure screw nut, the cable is fixed, and then tighten the fastening screws around the cover, the indicator of waterproof and dustproof;
    4, upstream and downstream piping and instrument instrument should be the same caliber, flange or threaded connection should be matched with the flange and screw instrument, instrument upstream straight pipe length should be 5 times the nominal diameter of the instrument, the downstream straight pipe length not less than H2 250mm;
    5, the instrument measuring mechanism is magnetic transfer, so in order to ensure the performance of rotameter installed around the sensor, at least 100mm, the existence of ferromagnetic material is not allowed;
    Before installation, 6 meter pipeline sweeping shall be carried out, to prevent the retention of ferromagnetic material in the pipeline is attached to the instrument, the instrument performance, or even damage the instrument. If unavoidable, should be installed at the entrance of the magnetic filter instrument;
    7, the measurement of gas meter is calibrated in special pressure, if the gas meter in the outlet directly discharged into the atmosphere, will produce a pressure drop in the float, and lead to data distortion.If this is the working conditions, should install a valve in the instrument exports, so that the required flow value is set, when the float to maintain calibration pressure, gas expansion in valve;
    The installation form 8, instrument installation for vertical and horizontal installation, if it is installed in a vertical straight form, should ensure that the instrument is better than 1%; if the level of installation form should ensure that the level of less than 1% meters;
    9, installed in the pipeline in the instrument should not be subject to stresses, the instrument should be a proper entrance and exit pipe support, can make the instrument in the minimum stress state;
    10, product and computer related to calibration and other operations, must be in a safe place.
    Eight, installation notes:
    In 1, before the first installation flowmeter in the pipeline, dirt, impurities such as welding blowing clean, if the liquid contained solid impurities, should the corresponding filter is installed between the valve and the straight pipe.
    2, shielding and grounding system: in order to ensure ideal electromagnetic compatibility, shielding and grounding is very important, especially the bus cable, cable as far as possible to make continuous shielding.
    3, the safety zone, according to the requirements of conventional cable grounding as much as possible.
    4, in dangerous places, to ensure that the whole range of bus safe places and dangerous enough, all the potential, in this case, multiple ground is very superior.

    Nine, choose the flow table:

    Decay path

    Common path

    measuring range

    The maximum pressure loss

    DN (mm)

    DN (mm)

    The air (m3/h)

    Water (L/h)

    The air (KPa)

    Water (KPa)

    Fifteen

    Fifteen

    0.05-0.5

    1.6-16

    Seven

    Six point four

    0.07-0.7

    2.5-25

    Seven point one

    Six point five

    0.11-1.1

    4.0-40

    Seven point two

    Six point five

    0.18-1.8

    6.0-60

    Seven point three

    Six point six

    0.28-2.8

    10-100

    Seven point five

    Six point six

    0.40-4.0

    16-160

    Eight

    Six point eight

    0.70-7.0

    25-250

    Ten point eight

    Seven point two

    1.00-10

    40-400

    Ten

    Eight point six

    1.60-16

    60-600

    Fourteen

    Eleven point one

    Twenty-five


    80-800


    Fourteen

    Twenty-five

    3.00-30

    100-1000

    Seven point seven

    Seven

    4.50-45

    160-1600

    Eight point eight

    Eight

    7.00-70

    250-2500

    Twelve

    Ten point eight

    11-110

    400-4000

    Nineteen

    Fifteen point eight

    Fifty


    500-5000


    Sixteen point two

    Fifty

    18-180

    600-6000

    Eight point six

    Eight point one

    25-250

    1000-10000

    Ten point four

    Eleven

    Eighty

    40-400

    1600-16000

    Twelve point six

    Seventeen

    Eighty

    50-500

    2000-20000

    Fifteen point five

    Six point three

    One hundred

    75-750

    2500-25000

    Seventeen Point Two

    Eight point one

    One hundred

    100-1000

    4000-40000


    Nine point five

    One hundred and twenty-five

    160-1600

    6000-60000


    Ten

    One hundred and twenty-five

    250-2500

    8000-80000


    Sixteen

    One hundred and fifty


    10000-100000


    Twenty


    300-3000

    12-120m3/h


    Twenty


    400-4000

    15-150m3/h


    Twenty-five

    Ten, selection table:

    Specification code

    Explain

    The name of the factory

    ZW

    Zhongwei control

    Instrument types

    LZZW

    The micro flow of metal tube flowmeter

    structural style

    A

    In the next

    B

    Progress under the

    C

    Side measure

    D

    The bottom side

    R

    Right into the left out

    L

    The left into the right

    K

    Clamp connection (sanitary polishing tube)

    W

    Threaded connection

    Size code

    - XXX

    DN15 said: 15 cases

    The measuring tube material

    R1

    1Cr18Ni9Ti, 304

    R0

    0cr18Ni12M02Ti, 316

    RL

    316L

    RP

    PTFE (PTFE)

    Ni

    Hastelloy (Hastelloy C)

    Ti

    TitHXium (Tai Hejin)

    Media type

    L

    liquid

    G

    Gas

    working temperature

    H1

    - 40 -100 OC

    H2

    - 80 -250 OC

    Rated pressure

    One point six

    1.6MPa

    Four

    4.0MPa

    Sixteen

    16MPa

    Twenty-five

    25MPa

    Additional structure

    O

    nothing

    Z

    Damping type

    T

    Jacket

    H

    High temperature type

    Y

    High pressure type

    Form indicator

    M1

    Local indication

    M2

    Intelligent

    M3

    Intelligent (explosion)

    The output signal in the communication

    N

    nothing

    A

    4-20mA

    H

    4-20mA+Hart

    Alarm controller

    K1

    The upper limit alarm

    K2

    Alarm

    K12

    The upper and lower limit alarm

    Explosion proof type

    E1

    Intrinsically safe EXiaCT5 square casing

    E2

    Flameproof shell EXiibt4 Park

    Power supply

    Zero

    24VDC (optional)

    Two

    220VAC

    Three

    The 3.6 volt lithium battery